Naso‐gastric Naso‐duodenal Naso‐jejunal NG ND NJ Bedside Mouth (oro‐) Oro‐gastric OG Bedside Long‐term feeding devices Percutaneous Stomach Jejunum Gastrica Gastro‐jejunostomy Jejunostomy G‐tube GJ J‐tube Surgically or endoscopically Fluoroscopically Fluoroscopically

      a Commonly, these tubes are called “PEG tubes”; however, a percutaneous endoscopic gastrostomy is a procedure and not a specific tube.

      

      Temporary Feeding Devices

      Naso‐gastric (NG), naso‐duodenal (ND), NJ, and oro‐gastric (OG) feeding tubes are used for short‐term enteral feeding, defined as that less than 12 weeks. OG tubes are reserved for patients in the intensive care unit and rarely seen in the emergency department. Temporary feeding tubes are typically constructed from polyurethane or silicone‐based polymers, both of which are flexible, reasonably durable, minimally reactive biologically, and, for most, immunologically inert. Polyurethane tubing has the added benefit of being made with a water‐activated lubricant to ease insertion and increased durability. Depending on the manufacturer, NG, NJ, ND, and OG tubing may come with weighted tips, radiopaque indicators, stylets, and/or magnets to help with placement. Common pediatric tube size for feeding is a 6–8 Fr and for adults a 12–14 Fr.

      Gastric Decompression Devices

      Similar to the temporary feeding devices, there are NG and OG tubes used for decompression and lavage. These devices are larger than feeding tubes: 8–10 Fr for children and 12–14 Fr for adults. Decompression devices are divided into single and double lumen tubes. Single lumen tubes, such as the Levin tube, are used more frequently in the emergency department or intensive care unit for intermittent decompression. Single lumen tubes should not be placed on continuous suction because they can adhere to the stomach wall and cause tissue damage. A double lumen tube is the preferred decompression device because it has both a large lumen for suction or irrigation and a small lumen (typically blue in color) that vents the large lumen. This small lumen serves as a pop‐off valve for the device to prevent excessive suctioning. There are two common types of double lumen tubes: Salem sumps and Replogle tubes. Salem sumps are preferred in an emergency setting because they have several suction holes along the side of the tubing for rapid efficient suctioning, whereas Replogle tubes have suction holes only at the most distal end of the tube.

      Long‐term Feeding Devices

      G‐tubes are used for long‐term or permanent enteric feeding. G‐tubes are divided into standard adjustable length tubes and low‐profile (i.e. button) tubes. When caring for a patient with a long‐term feeding device, it is imperative that you know the type of tube the patient has, how the tube was placed, and how to use the tube in order to adequately care for your patient.

      There are advantages and disadvantages to balloon and nonballoon G‐tubes. The benefit of balloon G‐tubes is that they can be replaced at home; however, they are not as well tolerated as nonballoon retention devices because of the size of the balloon. Furthermore, balloon retention devices need to be changed more frequently than nonballoon G‐tubes (every three months compared to every six months, respectively). The main disadvantage of nonballoon G‐tubes is that every tube change has to be done by a medical professional.

      Standard G‐Tube

      Standard G‐tubes are adjustable length tubes that have an external bolster, which sits on the skin and can be moved up and down to adequately secure the tube in a patient of any size (Figure 1.1 a and b). These are particularly helpful in patients with increased soft tissue or in patients with a projected weight gain where a low‐profile tube with a fixed shaft length may not fit properly. Standard G‐tubes can be placed surgically or endoscopically.

Schematic illustration of the standard G-tubes and low-profile G-tubes. (a) Standard GJ tube with three ports: balloon port, jejunal port, and gastric port. (b) Standard G-tube with three ports: medication port, gastric port, and balloon port. (c) Low-profile tubes with both nonballoon and balloon retention devices.

      Standard G‐tubes are placed endoscopically by using the percutaneous endoscopic gastrostomy(PEG) technique. Of note, the term “PEG” is used inaccurately in medical vernacular to refer to all kinds of G‐tubes, but a PEG is actually the procedure and not a type of tube. During a PEG procedure, an endoscope is used to transilluminate the stomach and identify the stoma site. A needle is then inserted through the skin into the stomach with a guidewire that is pulled up through the esophagus and out of the mouth. This guidewire is then used to guide the G‐tube into the stomach. A small incision is made, and the G‐tube is pulled through the stomach and abdominal walls and secured in place by the internal