id="ulink_d75dfa7f-bfa6-5e43-aab9-9d91ba85d0c4">The LVAD patient in distress might be having an issue with the device, an exacerbation of the underlying cardiac disease, or an unrelated medical event. The initial EMS assessment should be to determine if the issue is LVAD‐related or not (Figure 11.3). If the event does not seem to be LVAD‐related, then local protocols or medical oversight should be consulted for further guidance. The next step is to determine the type of LVAD involved. The patient and caregiver should have device information available. This information should include whether the patient can receive electrical therapy and whether or not CPR can be performed. Obviously, these questions need immediate answers [11, 21].

Schematic illustration of emergency assessment of a patient with an LVAD.

      Patients with evidence of hemodynamic compromise or hypoperfusion should have large‐bore IV access, and be volume resuscitated. Vasopressors are not generally a good initial therapy, as many problems are volume‐related, and vasopressors will increase afterload, which can worsen pump flow [11].

      Arrhythmias should only be treated if they are symptomatic. An LVAD patient with full left ventricle support may be able to tolerate ventricular tachycardia or fibrillation. If the arrhythmia requires treatment, the usual therapies can be used for rate control and rhythm conversion. The patient can also receive electrical therapy [11]. Defibrillator pads should not be placed over the device. Some devices may require that the system controller cables be disconnected prior to defibrillation to prevent damage to the electronics. The patient should also be examined for the presence of an implantable cardioverter defibrillator (ICD), which should provide the appropriate treatment in the event of ventricular arrhythmia [22].

      The decision of when to perform CPR can be a major conundrum in treating these patients. Knowledge of the device type and function is crucial. Patients with first‐generation LVADs producing pulsatile flow should not receive chest compressions. Instead the hand pump should be used. Second‐generation and later continuous‐flow devices will not have a hand pump. Chest compressions carry the risk that they may dislodge the device, resulting in exsanguination and death. On the other hand, if the LVAD is not pumping, the underlying left ventricle will not have the ability to maintain perfusion of organ systems. Patient survival is not likely. Lack of compressions may also result in a thrombus formation in the pump, resulting in obstruction to pump flow, and potential downstream embolic events. Awareness by EMS clinicians of the patient’s advanced directives regarding resuscitation may be important, as these patients have chronic severe disease. Ideally, device information, patient wishes and treatment plans, and contact information should be prepared prior to initial discharge from the hospital. In the event of an EMS contact with a patient who is hypoperfused and has a nonfunctioning pump, an attempt may be made to contact the LVAD coordinator for further recommendations. If the coordinator cannot be reached, and the patient is to be resuscitated, compressions should be started and transport initiated per local protocols [11].

      The LVAD patient should be transported to the hospital that placed the device if possible. These hospitals are usually tertiary care centers and should be capable of managing not only LVAD complications but also other issues, such as stroke or GI bleeding. If there are distance issues, air medical transport should be considered. This can shorten transit time and also provide critical care services. Regardless of transport mode, the LVAD patient should be transported with all device equipment, batteries, controllers, documentation, and caregivers (if possible).

      EMS personnel may also encounter patients in the field with short‐term electrical support devices, such as wearable cardioverter defibrillators (WCDs), or long‐term, implanted cardiac devices, such as implantable cardioverter defibrillators (ICDs) and pacemakers. The approach to a patient with an electrical cardiac support device who is suffering from a medical condition is to determine if the problem results from the device or the underlying medical condition. While these devices are generally safe and effective, thousands of patients have been affected by pacemaker and ICD malfunctions [23].

      Pacemakers

      Cardiac pacemakers are implanted in patients who suffer from bradycardic dysrhythmias. If the patient’s intrinsic rhythm falls below a set target, the pacemaker will provide an electrical stimulus to the myocardium.

Скачать книгу

AOO Atrial pace; no sense, no inhibitions
AAI Atrial pace; atrial sense, inhibited by atrial beat
VOO Ventricular pace; no sense, no inhibitions
VVI Ventricular pace; ventricular sense, inhibited by ventricular beat