Process Gas Chromatographs. Tony Waters

Чтение книги онлайн.

Читать онлайн книгу Process Gas Chromatographs - Tony Waters страница 24

Process Gas Chromatographs - Tony Waters

Скачать книгу

of the carrier gas. The solubility of the molecules is again assumed to be 50 %. The carrier movement represents the distance moved by carrier gas molecules that were present at the instant of sample injection.

      Each colored trace shows the distribution of molecules and the corresponding peak shape obtained for a different number of equilibria (N). In practice, the narrower peaks would be much higher: the scale of the vertical axis is not the same for each trace. When a column operates under optimum conditions, a peak experiences more equilibria as it passes through the same length of column, resulting in narrower peaks that are easier to separate. Graph depicts an effect of having more equilibria. Each colored trace represents the distribution of molecules and the corresponding peak shape obtained for a different number of equilibria. The scale of the vertical axis is not the same for each trace, in which the narrower peaks would be much higher. When a column operates under optimum conditions, a peak experiences more equilibria as it passes through the same length of column, resulting in narrower peaks that are easier to separate.

      The gold curve in Figure 2.9 is a smoothed version of the 1:4:6:4:1 distribution we obtained with five equilibria. This embryonic peak would appear in the first two millimeters of a regular packed column and would take less than one fourth of a second to form.

      Continuing the jerky mechanism for more equilibria would be tedious, but luckily it can be done mathematically. Figure 2.9 also displays the curves for 50, 500, and 5,000 equilibria, plotted as if they occurred along the same length of column. They look just like chromatogram peaks! And that's exactly what they are. You have just witnessed how the standard peak shape forms.

      The plate number is a useful parameter for evaluating column performance. As is evident from Figure 2.9, anything that increases the plate number must be a good thing, because it reduces peak width and narrower peaks provide better separations.

      Figure 2.9 contains a lot of information. Let's pause for a moment and see what it can tell us.

      Identical molecules – different behavior

      Notice that the propane molecules do not all spend the same time in the column, even though they are identical to each other. They enter the detector at different times.

      The variable arrival time of identical molecules at the detector is the root cause of chromatogram peak shape.

      All peaks are symmetrical

      The statistical bell‐shaped curve is named for Carl Friedrich Gauss, the acclaimed German mathematician who derived its mathematical equation. The two major theories of chromatography both deduce that all the peaks in the column should be Gaussian in shape.

      Thus, our ideal peaks in Figure 2.9 are perfectly symmetrical. Unfortunately, real peaks are not perfect, and they may be slightly or grossly asymmetric due to other mechanisms occurring in the columns or in the instrument itself.

      More equilibria – narrower peaks

      On a given column, the peaks always get narrower as the number of equilibria increases. This is a very important observation because narrow peaks are highly desirable. It's so much easier to separate narrow peaks from each other than to separate wide peaks from each other. Therefore, to ensure that the peak widths are as narrow as possible, we adjust the column operating conditions − particularly the carrier gas flow rate − to generate the maximum number of equilibria.

      In practice, each peak exhibits a different plate number. We can calculate the plate number of a selected peak from chromatogram measurements of its width and retention time. This calculation is used to evaluate column performance and to optimize column operating conditions. We'll look at it later.

      It follows that a certain length of column must be necessary to generate each equilibrium. You can calculate it if you wish: just divide the column length by the plate number. In most columns the answer will be about 0.5 mm. We call this the plate height (H). It's a very important parameter for optimizing the performance of a column.

      More equilibria – taller peaks

      The apex of the peak gets higher when the plate number increases. It actually gets very much higher because the peak area remains constant (the vertical axis in Figure 2.9 is not drawn to scale). A higher peak in the graph means there are more molecules in the detector and a larger detector output signal, making the peak easier to measure. The chromatogram baseline always suffers from a little detector signal noise, so a higher peak has a better signal‐to‐noise ratio.

      Retention at the apex

      More equilibria – same retention time

      In Figure 2.9, the retention time of the propane peak is always the same, regardless of the plate number. The plate number of a column affects only the peak width; it has no effect on the peak retention time.

      In Figure 2.9, the peak apex has traveled along the column exactly half the distance that the carrier gas has traveled. This remains true during all of its journey through the column. When the carrier gas reaches the end of the column, as depicted, the propane peak has reached the mid‐point. At that time, it's only halfway through the column.