Process Gas Chromatographs. Tony Waters

Чтение книги онлайн.

Читать онлайн книгу Process Gas Chromatographs - Tony Waters страница 28

Process Gas Chromatographs - Tony Waters

Скачать книгу

convenient values for the component solubilities are just simple examples assumed for discussion purposes, but every substance has a real solubility in a given liquid that depends only on the temperature and pressure.

      A challenge question

      Imagine that the effluent from the column in Figure 3.1 flows into a detector and the detector signal is recorded in the form of a chromatogram. For this exercise, assume Figure 3.1 shows the correct position of the four peaks in the column at three minutes after sample injection.

      Now predict what the chromatogram will look like after all the peaks have exited the column.

       It's worth your time to stop for a moment and try to do this; you will learn a lot from the exercise. Be careful; it's more difficult than it looks!

      Based on the information given in Figure 3.1, draw the chromatogram you would expect to see after all four peaks have passed through the detector. Your chromatogram should show the four peaks at their correct retention times. The injection marker is time‐zero on your chromatogram, but you must decide the time scale. In this exercise, don't worry about peak widths. Graph depicts a blank chromatogram. Draw the chromatogram that would expect to see after all four peaks have passed through the detector. The chromatogram should show the four peaks at their correct retention times. The injection marker is time-zero on your chromatogram, but you must decide the time scale.

      How confident are you of your chromatogram? People rarely get it right at first attempt. The most common mistakes are:

       Inserting all four peaks between 0 and 3 mins on the chromatogram

       Assuming all four peaks are equally spaced

       Assuming three of the peaks are equally spaced

       Assuming any peak has a fractional retention time (none do)

      Want to try again? We'll reveal the correct answer later.

      Significance of the air peak

      The air peak is a valuable indicator of column performance. Since air doesn't dissolve in column liquids to any significant extent, the air peak remains in the gas phase all the way through the column – traveling at full gas velocity. We call it an unretained peak. Its position on the chromatogram indicates the elapsed time for the carrier gas to travel from one end of the column to the other end.

      If using a detector that doesn't respond to air, another unretained peak can act as a surrogate. For instance, methane often serves as an adequate “air peak” on a flame ionization detector.

      Since we know the column length (L), the air peak retention time (tM) allows us to calculate the average carrier gas velocity (uM) in m/s:

      We don't need this information now, but Equation 3.1 is used in more advanced work to optimize the carrier gas flow rate.

      The air peak location is good to know when troubleshooting because component molecules can't travel faster than the carrier gas, so no peaks from the current sample injection can elute before the air peak.

      But the air peak has a much more significant role because it shows how long the carrier gas took to pass through the column. Recall that component molecules can move along the column only when they are in the carrier gas. So, to get through the column, all component molecules must spend the same time in the carrier gas as the air peak does.

      As a way of emphasizing the point, imagine that the column had no liquid inside; where would the peaks be on the chromatogram?

      Yes, of course, all the peaks would elute together with the air peak. With no liquid present, no separation can occur. The air peak reveals the time that component molecules spend in the gas phase, and it's the same for every component.

      As an aside, realize from this imagined experiment that the time a peak spends in the liquid phase is proportional to the amount of liquid phase present in the column. In packed columns, the liquid film can slowly evaporate into the carrier gas, thereby, over time, reducing the observed peak retention times.

      An air peak tells us how long the carrier gas takes to get through the column, which is also the time that each component spends in the gas phase. The additional retention time for each peak is the time that peak spends in the liquid phase. Graph depicts the significance of an Air Peak. An air peak tells us how long the carrier gas takes to get through the column, which is also the time that each component spends in the gas phase. The additional retention time for each peak is the time that peak spends in the liquid phase.

      Notice that the air peak neatly divides the chromatogram into two time zones and makes them easy to measure:

       From injection to the air peak is the time that all components spend in the gas phase. This time is often called the “dead time” because it doesn't contribute to separation. The formal name for the time in the gas phase