the “lung line” is now too bright, lacking the contrast you need for Vet BLUE. Conversely, moving from Vet BLUE to AFAST or TFAST, you are now too dark (undergained) because the “lung line” imaging did not need as much gain as the soft tissue of abdominal and cardiac structures.

       Check your depth. Depending on the area of interest and the size of your patient, paying attention to the depth will help you to get where you want to go more quickly (see Figure 5.13). Generally, it’s best to start with too much depth to gain an overview of what you are imaging, and to survey for recognizable landmarks prior to magnifying (decreasing the gain). And time is valuable on many fronts, including the patience of your patient and its tolerance of your physical and chemical restraint.Figure 5.14. Splenic mass and estimating size from the centimeter scale. These images were captured during a POCUS spleen following AFAST. In (A) is the spleen (*) and arrows showing its borders. The spleen is sonographically recognized by its hyperechoic (bright white) capsule, and its splenic vessels splitting the capsule (not shown). In (B) the mass detected is eyeballed at approximately 4 × 4 cm using the centimeter scale. The mass appears large when in fact it is quite small. The finding is important and concerning because the mass deforms the capsule of the spleen.Source: Courtesy of Dr Gregory Lisciandro, Hill Country Veterinary Specialists and FASTVet.com, Spicewood, TX.Figure 5.15. An algorithm for trouble shooting image acquisition with some clinical pearls and descriptors.Source: Courtesy of Dr Gregory Lisciandro, Hill Country Veterinary Specialists and FASTVet.com, Spicewood, TX.Figure 5.16. Time gain compensation (sliders) and the overall gain (wheel). Shown in (A) is an ultrasound machine with its keyboard and screen. The time gain compensation (TGC) is in the upper right on this keyboard. These TGC sliders can be intimidating but gain adjusts your gray scale when in B‐mode. The sliders represent the different levels from the near‐field to the far‐field. In general, they should look like the circled upper two images in (B) and not like the lower two images (marked with an "X"). In (C) is the overall gain (circled wheel with arrow pointing to it) that may be turned clockwise and counterclockwise that alters the overall picture. If your sliders are adjusted as in (B) then you should mostly only be using the overall gain thereafter. If the image still doesn't look good then go through the trouble‐shooting algorithm (see Figure 5.15).Source: Courtesy of Dr Gregory Lisciandro, Hill Country Veterinary Specialists and FASTVet.com, Spicewood, TX.

       Check the location and numbers of your focus cursor(s). Generally, a single focus cursor is best that is centered on the screen or directly across your area of interest.

       Check your probe–skin contact (coupling). Is the probe head directly on skin? Do you have ample acoustic coupling medium?

       Check your probe hand placement and beam direction. Make sure that you are in the right region, using external landmarks as well as considering the direction and path of your ultrasound beam (see Figures 5.10 and 5.11).

       Default back to original settings. If you still cannot figure out why the image looks so wrong, push the B‐mode button first and then go to presets and change to a different preset or return to the original preset, as many machines will default back to the original settings. The reality is sometimes you have no idea what you did to make everything look so bad.

      See Chapter 4 for additional explanation and examples.

       Not Knowing How to Freeze and Roll the Cine Ball

Image described by caption.

      Source: Courtesy of Dr Gregory Lisciandro, Hill Country Veterinary Specialists and FASTVet.com, Spicewood, TX.

       Lacking Awareness of Most Common Locations for Artifacts

      Saving images is important not only for archiving patient information but for learning by reviewing studies and being able to compare to other modalities, future studies, and clinical course and outcome. For learning artifacts and how your machine interprets them, you can go through studies playing the “artifact game.” For example, look at the AFAST views for mirror image artifact at the DH view on the pleural cavity side of the diaphragm, the CC view and the other side of the abdominal wall, and TFAST echo views for mirror image artifact on the far side of the pericardium. This can also be done with other artifacts in their most common locations, such as edge shadowing off the curved surface of the stomach wall, renal cortex, urinary bladder, etc.

       Mirror Image

      This most commonly occurs at the diaphragm but the most problematic is the mirroring of the gallbladder into the thorax, especially when it’s a partial mirror image. In this case, the gallbladder can mimic pleural and pericardial effusion to the hasty sonographer.

      Other places for mirror image, where generally a strong air–soft tissue interface exists, are on the far side of the heart when aerated lung is in direct opposition to the pericardium, and the urinary bladder against the colon or thin body wall. See Chapter 3 for more explanation and examples.

      